《中国康复理论与实践》 ›› 2014, Vol. 20 ›› Issue (4): 371-373.

• 论文 • 上一篇    下一篇

肌肉评定和训练系统测定健康成人上肢远端肌力的重测信度研究

陈颖,李放,徐秀林   

  1. 1.复旦大学附属华山医院北院康复医学科,上海市201907;2.复旦大学附属华山医院康复医学科,上海市200040;3.上海理工大学,上海市200093。
  • 收稿日期:2013-04-07 修回日期:2013-10-30 出版日期:2014-04-25 发布日期:2014-04-25
  • 通讯作者: 李放

Test-retest Reliability of Distal Upper Extremity Muscle Strength by Muscle Evaluation and Treatment System on Healthy Adults

CHEN Ying, LI Fang, XU Xiu-lin.   

  1. North Institute of Huashan Hospital Affiliated to Fudan University, Shanghai 201907, China
  • Received:2013-04-07 Revised:2013-10-30 Published:2014-04-25 Online:2014-04-25

摘要: 目的评估肌肉评定和训练系统对健康成人上肢远端肌力的重测信度。方法采用上海理工大学和华山医院康复科合作研发的肌肉评定和训练系统对30 名健康成人进行测试,由1 名评估者分别对受试者双侧屈腕肌群、指浅屈肌、指深屈肌、指伸肌、伸腕肌群进行肌力测试,并测定每组肌群的峰力矩、峰力矩体重比、最大单次做功量、最大单次做功量体重比、平均力矩、平均功率各项参数。测试共进行2 次,分别在第1 天和1 周后进行。结果肌肉评定和训练系统对健康成人的指浅屈肌峰力矩(ICC: 0.814~0.870)、指深屈肌峰力矩(ICC: 0.811~0.848)、屈腕肌群峰力矩(ICC: 0.871~0.808)、伸腕肌群的峰力矩(ICC: 0.778~0.873)有好的重测信度;对健康成人的指伸肌峰力矩有中等至好的重测信度(ICC: 0.302~0.803)。此外,肌肉评定和训练系统对健康成人的各项肌群的平均功率有中等至好的重测信度。结论肌肉评定和训练系统在健康成人上肢远端肌力观测指标中的峰力矩和平均
功率有较好的重测信度。

关键词: 肌肉评定和训练系统, 重测信度, 上肢

Abstract: Objective To investigate the test-retest reliability of Muscle Evaluation and Treatment System on healthy adults. Methods 30 healthy adults participated in this study and were tested by Muscle Evaluation and Treatment System, designed by University of Shanghai for Science and Thechnology and Rehabilitation Department of Huashan Hospital, on muscle strength of both sides' flexor carpi muscles,musculus flexor digitorum sublimis, flexor digitorum profundus, extensor digitorum muscle, extensor carpi muscles. Peak torque, peak torque per weight, largest single work, largest single work per weight, average torque, average power were the main parameters. Each participant was tested 2 times, and the interval was 7 days. Results Almost perfect test-retest reliability of peak torque was performed on flexor carpi muscles, musculus flexor digitorum sublimis, flexor digitorum profundus, and extensor carpi muscles. Moderate to excellent test-retest reliability of average power were performed on extensor digitorum muscle. Conclusion Good test-retest reliability of Muscle Evaluation and Treatment System is proved in the study.

Key words: Muscle Evaluation and Treatment System, test-retest reliability, upper extremity